Turbidity
Reference:
Standard
Methods; 2130 A, 2130 B
Scope
and Application
Turbidity
of water is due to suspended solids such as clay, plankton, silt, finely divided
organic matter, microscopic organisms and similar materials. These solids will
deflect (or scatter) light as it passes through the sample. Turbidity is a measurement
of the scattered light as compared to the amount of light scattered by a standard.
The more light that is deflected the higher the turbidity of the sample. The
measuring device used in today’s laboratories is called a nephelometric meter.
This type of meter does not measure all of the deflected light, only that which
is deflected at a right angle (90° ) from the sample and light source. Turbidity
is read as nephelometric turbidity units (NTU).
Apparatus
- Turbidimeter: with a
readout device to indicate intensity of light scattered at 90° . Due to the
nature of the test, the meter must not allow any outside light to enter the
detector area. The turbidimeter must be free from drift and be able to detect
differences of 0.02 NTU or less for samples with turbidity of less than 1 NTU.
The instrument should detect in the range of 0 to 40 NTU.
Design
differences will cause variations in measured values for turbidity. To minimize
these differences, select an instrument that meets the following criteria:
- Light source: Tungsten-filament
lamp operated to meet U.S. Environmental Protection Agency NPDES requirements.
- Distance traversed by
incident light and scattered light within the sample tube should not exceed
10 cm.
- Detector: The angle
of the light acceptance detector must be centered at 90° to the incident light
path. The angle of the light acceptance detector must not exceed ± 30° from
the required 90° . The detector and filter system shall have a spectral peak
response between 400 and 600 nm (nanometers).
- Sample Tubes are made
of a clear colorless glass. Dust, dirt scratches and fingerprints deflect
light. Therefore, keep the tubes scrupulously clean at all times. Touch tubes
only around the top to avoid touching where light must pass. Throw away scratched
or etched tubes.
Reagents
- Turbid free water -
if the turbidity of the laboratory grade water is 0.05 NTU or higher,
pass the water through a membrane filter having precision-sized holes of 0.2 m
m. Rinse collecting flask twice with filtered water. Discard the next 200 mL
of filtered water, then start collecting filtered water to prepare standards.
Commercially prepared water can be substituted when its turbidity is lower
than what is available in the laboratory.
- Stock Turbidity Suspension
Solution 1 - Weigh 1.00 gram of hydrazine sulfate, (NH2)2·H2SO4.
Dissolve the 1.00 gram of hydrazine sulfate in a 100 mL volumetric
flask. Use turbid free water to fill to the 100 mL line on the volumetric
flask Prepare stock solution monthly.
Caution:
hydrazine sulfate is a known carcinogen: avoid inhalation, ingestion and contact
with skin.
- Stock turbidity Suspension
Solution 2 - Weigh 10.00 grams hexamethylenetetramine, C6H12N4.
Dissolve the 10.00 grams of hexamethylenetetramine in a 100 mL volumetric
flask. Use turbid free water to fill to the 100 mL mark on the volumetric
flask. Prepare stock solution monthly.
- 400 NTU Turbidity Solution
- Mix 5.0 mL each of Solution 1 and Solution 2 in a 100 mL volumetric
flask. Let stand for 24 hours at 25 ± 3° C, then fill to the mark with turbid
free water and mix.
- 40 NTU Standard Solution
(working solution) - Measure 10.00 mL of stock 400 NTU Turbidity
solution. Add the 10.00 mL of stock solution to a 100 mL volumetric
flask. Fill to the 100 mL mark with turbidity free water. Prepare this
solution daily.
- Alternate Standards
- Commercially prepared standards are available.
Storage
/ Preservation
The turbidity
test should be determined on the day the sample is taken. If this is not possible,
refrigerate the sample at 4° C for up to 24 hours. Remember to vigorously shake
all samples before examination.
Raw
Data Sheet Format
The following
must be recorded on the data sheet:
- Source identification
(source, name, and date(s) of collection)
- Final results with correct
units (reported to nearest NTU)
- Description of unusual
sample characteristics
- Replicates are to be
listed in an orderly cluster
Quality
Control Requirements
- If a pre-calibrated
scale is used, at least one standard must be analyzed in each range to check
the operation of the instrument
- A series of standards
should be used to calibrate the instrument in each range of interest
- Sample cells should
be free of contamination, etches, and scratches
Procedure
Calibration
Always
follow the manufacturer’s instructions for calibration of your particular meter.
If the instrument does not have a prepared calibration curve, make one by using
various values of turbidity standards. Plot turbidity reading vs. the standard
concentration to obtain curve.
Check the
accuracy of the instrument against bubble-free prepared standards, make adjustments
in readings according to manufacturer’s instructions. Run a minimum of one standard
for each range used during the test, making sure that the meter gives stable
readings in all sensitivity ranges used.
Sample
Analysis
Measurement
of turbidities less than 40 NTU.
- Thoroughly shake sample.
Wait until air bubbles disappear before pouring sample into turbidity tube.
When necessary, immerse turbidity tube in an ultrasonic bath for 1 to 2 seconds
to dislodge bubbles. Letting the sample stand for a period of time to allow
air bubbles to dissipate will also allow solids to settle thus changing the
characteristics of the sample being evaluated.
- Wipe outside of tube
to remove fingerprints, dust dirt, and water droplets. Place tube in turbidimeter.
- Read turbidity from
instrument direct reading scale or convert from calibration curve.
Measurement
of turbidities greater than 40 NTU
- Thoroughly shake sample.
Dilute samples using turbid free water until sample falls between 30 and 40
NTU.
- Wait until air bubbles
disappear before pouring sample into turbidity tube. When necessary, immerse
turbidity tube in an ultrasonic bath for 1 to 2 seconds to dislodge bubbles.
Letting the sample stand for a period of time to allow air bubbles to dissipate
will also allow solids to settle thus changing the characteristics of the
sample being evaluated.
- Wipe outside of tube
to remove fingerprints, dust, dirt, and water droplets. Place tube in turbidimeter.
- Read turbidity from
instrument direct reading scale or convert from calibration curve.
- Calculate the turbidity
of the original sample using turbidity of diluted sample and the dilution
factor.
Calculations

Where:

Example:
If 40 mL
of sample is diluted to 200 mL with 160 mL of turbid free water and
the diluted sample showed a turbidity of 30 NTU, what is the turbidity?


The turbidity
of the original sample is 150 NTU.
Report
turbidity readings as follows:
|
Turbidity
Range NTU
|
Report
to the Nearest NTU
|
|
0-1
|
0.5
|
|
1
– 10
|
0.1
|
|
10
– 40
|
1
|
|
40
– 100
|
5
|
|
100
– 400
|
10
|
|
400
– 1000
|
50
|
|
>
1000
|
100
|
Turbidity
Analysis; the Tips for Techs version
- Collect sample
- Prepare raw data sheet
- Calibrate the turbidimeter
- Add sample to turbidity
tube
- Read turbidity of sample
from turbidimeter and record result