Standard Methods; 4500-H+ A, 4500-H+ B
Although this is one of the simplest tests, measurement of pH is one of the most important and most frequently used tests in wastewater analysis. The definition of pH is the negative logarithm of the hydrogen ion activity. In most instances that are dealt with in wastewater treatment, it is safe to assume that the activity coefficient will be 1, therefore the pH will be the negative logarithm of the hydrogen ion concentration.
pH is shorthand for expressing the negative logarithm of the hydrogen ion concentration, [H+], and is the measurement of hydrogen [H+] and hydroxyl [OH-] ions that are present. A sample having a equal number of each is considered neutral, or pH 7.00. When a sample has a higher amount of [H+] than [OH-], it is considered acidic - the pH is lower than 7.00, or between 0.00 and 6.99. Some examples of acids are nitric acid (HNO3) and sulfuric acid (H2SO4). When samples have a lower amount of [H+] than [OH-], the sample is basic - the pH is higher than 7.00 or between 7.01 and 14.00. Some examples of bases are potassium hydroxide (KOH) and sodium hydroxide (NaOH).
The pH notation of 0.00 - 14.00 was first introduced by a Danish biochemist Søren Sørensen. Before this, acid and basic solutions were noted in concentrations of [H+] and [OH-] in a solution. For example 0.010 M HNO3 (M is shorthand for the molarity of a solution) or 0.010 M NaOH. These are frequently written in scientific notation as 1.0 X 10-2, meaning that the one is two (2) places behind (-) the decimal point.
Acidic pH relates to the molarity by the following equation:
pH = - log [H+]
If a solution of hydrochloric acid is 1.0 X 10-2 M, then the pH of the solution is:
pH = - log [1.0 X 10-2]
pH = -(-2) or 2
Basic solution pH also relates to the molarity using a different equation:
pH = 14 - (-log [OH-])
If a sodium hydroxide solution is 1.52 X 10-4 M, then the pH of the solution is:
pH = 14 - (-log [1.52 X 10-4] )
pH = 14 - (-(-3.82))
pH = 14 - (3.82)
pH = 10.18
A rise in pH from 6.00 to 7.00 is really a ten-fold change in the potential of hydrogen. The molarity of the solution at 6.00 is 10-6 M hydrogen, where the 7.00 solution is 10-7 M hydrogen.
Taking a pH reading today has been greatly simplified by the use of a pH electrode and meter. The pH meter measures milliVolts (mV) and changes this into a pH reading by using the Nernst equation:
![]()
Where:
E = the potential in milliVolts developed between the sensing and reference electrodes
E0 = varies with the choice of electrodes, temperature, and atmospheric pressure.
R = gas constant = 8.31451*
T = Temperature in degrees Kelvin (Kelvin = 273.15 + ° C)
n = charge on the ion
F = Faraday constant = 96,485*
ai = the activity of the ion to which the electrode is responding
*These constants are related in terms of units, the discussion of which is beyond the scope of this manual.
The Nernst equation shows that it is very important that the temperature be taken into consideration when measuring the pH of a sample. This is often accomplished by using automatic temperature compensation (ATC). ATC is either built into a pH electrode or is a separate electrode placed into the sample along with the pH electrode. Temperature can also be compensated for manually if ATC options are not available. The change in pH values of some common buffers as related to temperature is listed in Table 1.
Table 1: Temperature versus pH
|
Buffer Value |
|||
|
Temperature (° C) |
4.01 |
7.00 |
10.00 |
|
0 |
4.00 |
7.14 |
10.30 |
|
5 |
4.00 |
7.10 |
10.23 |
|
10 |
4.00 |
7.07 |
10.17 |
|
15 |
4.00 |
7.04 |
10.11 |
|
20 |
4.00 |
7.02 |
10.05 |
|
25 |
4.01 |
7.00 |
10.00 |
|
30 |
4.01 |
6.99 |
9.96 |
|
35 |
4.02 |
6.98 |
9.92 |
|
40 |
4.03 |
6.98 |
9.88 |
|
45 |
4.05 |
6.98 |
9.85 |
|
50 |
4.06 |
6.98 |
9.82 |
|
55 |
4.07 |
6.98 |
9.79 |
|
60 |
4.09 |
6.99 |
9.76 |
1. Standardized Calibration Buffers at pH = 4.00, pH = 7.00, and pH = 10.00 are available commercially or can be made in-house with buffer tablets, buffer powders or reagent chemicals. The laboratory practices committee recommends that buffer solutions be purchased. It is generally recommended that properly stored stock buffers be used within the three month period after they are prepared.
2. Storage Solution for the electrode(s). Note that the electrode in the different probes require different storage solutions. Never store a probe in laboratory grade water. The electrolyte leaches into the water, shortening the life of the electrode. The solution should be slightly acidic to allow contaminants to be replaced by hydrogen ions.
3. Distilled water or other laboratory grade water. For rinsing off probe between samples or making standard buffers. Do not use this for storage of the probe.
There is no holding time or preservative for pH samples. Run sample as soon as possible after collection of the grab sample. Samples taken for regulatory purposes must be grabs.
Recorded data format
The following must be recorded on the data sheet:
The pH meter must be calibrated prior to analysis of each set of samples. A set is defined as any group of samples that can be analyzed consecutively without interruption.
Use a two-point calibration to calibrate pH meters. (For most analyses use pH 7 buffer as the isopotential or starting point.)
The pH meter should be calibrated across the range where the pH of the sample is expected to fall. It is acceptable to measure pH values up to 8.50 when the pH meter has been calibrated over the range of 7 to 4. Likewise, it is acceptable to measure pH values down to 5.50 when the pH meter has been calibrated over the range of 7 to 10.
Record all maintenance performed on the probe, such as adding filling solution or cleaning the junction.
Run duplicate analysis on one sample, the best choice being obtaining a second sample at the time of sampling.
If the sample pH falls outside the acceptable range for the calibration, the pH meter should be recalibrated to encompass the appropriate range.
The calibration of the probe and meter can be checked using a third buffer solution. Record the pH reading and buffer value on a bench sheet. For example – 7.00 and 10.00 were used to calibrate the meter. Use a buffer with value between 7.00 and 10.00 to check meter readings, or use the 4.00 standard buffer.
Recoveries of pH check samples should be between ± 2 standard deviations.
Recoveries of pH check samples must be between ± 3 standard deviations, or the pH meter calibration procedure will need to be repeated before proceeding with the analysis of samples.
Recoveries should be reviewed on a yearly basis and any changes documented in the check sample logbook.
Always follow the manufacturer’s instructions for pH meter calibration. If instructions are not available, use the following steps. It should be noted that individual instruments might require additional steps to be properly calibrated.
The following steps are general in nature to cover all the equipment generally used in laboratories.
The pH meter gives direct reading of pH. No calculations are needed.
Goals for a New Analyst Performing pH Analysis
pH meter, pH electrode, How to calibrate, General use and care of electrode
pH meter, pH electrode, pH buffer solutions
pH Analysis; the Tips for Techs version