%
Total Solids, % Volatile Total Solids
Reference:
Standard
Methods; 2540 A, 2540 B, 2540 E, 2540 G
Scope
and Applications
Total solids
are the sum of suspended solids and dissolved solids. Total solids analyses
are important for assessing wastewater treatment processes such as digester
efficiencies and sludge cake processing parameters. These tests are commonly
performed on sludge samples when the concentration of solids is greater than
10,000 mg/L or 1%, thus making them very difficult to filter.
The total
solids are composed of two components, volatile and fixed solids. The volatile
solids are organic compounds of animal or plant origin. Biological processes
can treat these. When volatile solids are referred to in this manual, we are
actually referring to the volatile total solids, or the portion of the total
solids that are volatile. The fixed solids are things such as sand, gravel,
and salt.
Apparatus
- Analytical balance -
capable of weighting to 0.1 mg
- Drying Oven for use
at 103 – 105°C
- Muffle Furnace for use
at 550°C
- Aluminum or ceramic
dish (permanently labeled)
- Crucible or evaporating
dish tongs
- Heat resistant gloves
- Desiccator and desiccant
that contains a color indicator for moisture content
- Cotton gloves
Reagents
The only
reagent that may be needed for % total solids analysis would be deionized
water, but even this is not necessary. This is because there is not a filtering
step and any material in or on the dish at the start of the analytical procedure
can be expected to remain constant through the process, (after the initial ashing
of the dish during dish preparation).
Storage
/ Preservation
Samples
may be stored in a plastic or glass container and kept for 7 days at 4° C. See
sample storage and preservation table.
Raw
Data Sheet Format
The following
must be recorded on the data sheet:
- Sample identification
(source, name, and date(s) of collection)
- Analyst(s)
- Raw data
- Final results with correct
units (reported to nearest 0.1%)
- Description of unusual
sample characteristics
- Replicates are to be
listed in an orderly cluster
Quality
Control Requirements
Replicate
analyses should yield RSD’s less than 5% for the entire data set.
Procedure
Dish
Preparation
- Pre-wash the aluminum
or ceramic dish.
- Place the dish in a
103° -105° C oven to dry. If volatile total solids are to be analyzed, move
the dry dishes into a 550° C muffle furnace for 15 minutes; if the volatile
portion does not need to be determined, place the dishes into a desiccator
to cool and skip step 3.
- Remove the dishes from
the muffle furnace and place on a heat resistant surface. The surface temperature
of the dish must be greater than 103° C when placed into the desiccator.
- Cool the dishes thoroughly
in a desiccator before use. Prepared dishes should be stored in a desiccator.
Sample
Analysis – % Total Solids
- Weigh the dish (at room
temperature) on an analytical balance. Use crucible tongs or wear lint free
cotton gloves to transfer the dish from the desiccator to the balance pan.
Handling the dish with your bare fingers may transfer oils and moisture from
the skin.
- Record the weight of
the dish.
- Thoroughly mix the sample
to be analyzed. Carefully measure the mass of sample transferred to the dish.
- Place the dish in the
oven to dry for one to two hours at 103° C. Depending on the mass of the sample
and the amount of water in the sample, it may take longer for the sample to
dry.
- When the sample is dry,
transfer the dried dish to a desiccator to cool. When the dish has cooled
sufficiently it should not feel warm to the touch on the inside of your forearm.
- Weigh the dried and
cooled dish on an analytical balance. Record the weight. If the sample is
not going to be used for regulatory purposes, it may be acceptable to use
this weight as the final dry weight.
- Return the dish to the
drying oven for another thirty minutes. Cool, desiccate, reweigh and record
weight. Repeat this procedure until the change in the weight of the residue
remains within 4% or less than 0.5 mg from one weighing to the next. (This
is referred to as constant weight.) Record the final weight on the benchsheet
and calculate the % total solids.
Calculations

Where:



Sample
Analysis – % Volatile Solids (or % Volatile Total Solids)
- Place the dish and residue
produced by %TS analysis in a muffle furnace at a temperature of 550° C for
15 to 30 minutes. The process is not complete until the heat has removed any
carbon or soot deposit (black glaze) on the dishes.
- Remove dish with ash
from muffle furnace, cooling partially, then placing in a desiccator until
cooled.
- Weigh cooled dish and
ash on an analytical balance and record weight. Be careful when opening the
desiccator. The cooled air (gas) inside the desiccator will have created a
vacuum, sudden opening of the desiccator may cause the loss of some of the
ash or fixed solids from the dish.
- Repeat ignition, cooling
and weighing until a constant weight is obtained or until weight loss is less
than 4% of the previous weight.
- Record the final ashed
weight on the benchsheet and calculate the % volatile total solids.
Calculations

Where:



Goals
for a New Analyst Performing %TS / %VTS Analyses
- Know safety procedures
for solids anlayses
- Know how to fill out
raw data sheets
- Know where to find equipment
/ supplies
- Understand equipment
used to perform total solids analyses
- Analytical scales
- Level of sensitivity
- How to tare weights
- Desiccator
- Muffle furnace
- Understand analysis
procedures used to perform total solids analyses
- Need for well mixed
samples
- Means for taking sample
aliquots
- Drying times
- Cooling times
- Know how to calculate
total and volatile total solids results
- Understand quality control
procedures
- Need for quality control
- Calculation of mean
and %RSD
- Know how to clean ceramic
or aluminum dishes
%
Total Solids Analysis; the Tips for Techs version
- Prepare ceramic or aluminum
dish (if not already done)
- Collect the sample
- Obtain dry weight of
ceramic or aluminum dish (tare weight)
- Transfer appropriate
volume or mass of sample to ceramic or aluminum dish
- Place ceramic or aluminum
dish and sample in oven to dry
- Move ceramic or aluminum
dish and sample residue to desiccator to cool
- Obtain dry weight (constant
if necessary) of ceramic or aluminum dish and sample residue
- Calculate total solids
- Place ceramic or aluminum
dish and sample residue into muffle furnace
- Move ceramic or aluminum
dish and sample residue to desiccator to cool
- Obtain ashed weight
of ceramic or aluminum dish and sample residue
- Calculate volatile total
solids
- Prepare ceramic or aluminum
dish for next or subsequent use